Hioki PW6001 benchtop power meters and power analyzers are best in class power measuring instruments for measuring single to three-phase lines with a high degree of precision and accuracy. The PW6001 is Hioki’s flagship power analyzer, featuring high accuracy, wide band, and high stability for measuring electrical power from DC to inverter frequencies, providing maximum of 12 channels* to support single- and three-phase inverter motor system measurements and next generation devices such as silicon-carbide (SiC) inverters. *When synchronizing two 6-channel models connected via optical link.
Measurement line type: Single-phase 2-wire, single-phase 3-wire, three-phase 3-wire, three-phase 4-wire
Number of input channels: Max. 6 channels; each input unit provides 1 channel for simultaneous voltage and current input (Voltage measurement unit: Photoisolated input, resistance voltage divider, Current measurement unit: Isolated input from current sensor)
Measurement items:
Voltage (U), current (I), active power (P), apparent power (S), reactive power (Q), power factor (λ), phase angle (φ), frequency (f), efficiency (η), loss (Loss), voltage ripple factor (Urf), current ripple factor (Irf), current integration (Ih), power integration (WP), voltage peak (Upk), current peak (Ipk)
Harmonic measurement: Harmonic active power, select calculation order from 2nd order to 100th order
Waveform recording: Voltage and current waveforms/ Motor pulse: Always 5 MS/s
Motor waveforms: Always 50 kS/s, 16 bits
Recording capacity: 1 Mword x ((voltage + current) x number of channels + motor waveforms)
Motor analysis (PW6001-11 to -16 only): Voltage, Torque, Rotation, Frequency, Slip or Motor output
Measurement range:
Voltage range: 6 V … 1500 V, 8 ranges
Current range (Probe 1): 400 mA … 1 kA (depends on current sensor)
Current range (Probe 2): 100 mA … 50 kA (depends on current sensor)
Power range: 2,40000 W … 4,50000 MW (depends on combination of voltage and current range)
Frequency range: 0,1 Hz … 2 MHz
Basic accuracy:
Voltage: ±0,02 % rdg. ±0,02 % f.s.
Current: ±0,02 % rdg. ±0,02 % f.s. + current sensor accuracy
Active power: ±0,02 % rdg. ±0,03 % f.s. + current sensor accuracy
Synchronization frequency range:
Power measurement: 0,1 Hz … 2 MHz
Harmonic measurement: 45 Hz … 66 Hz (IEC standard mode), 0,1 Hz … 300 kHz (Wideband mode)
Frequency band: DC, 0,1 Hz … 2 MHz
Data update rate:
Power measurement: 10 ms/ 50 ms/ 200 ms
Harmonic measurement: 200 ms (IEC standard mode), 50 ms (Wideband mode
Data save interval:
OFF, 10 msec … 500 msec, 1 sec … 30 sec, 1 minute … 60 minutes
User-selected from all measured values, including harmonic measured values,
Specified measured values can be saved in internal memory or USB flash drive
External interfaces: USB (memory), LAN, GP-IB, RS-232C (for communication/LR8410 link), External control, Synchronization control
Logger connectivity: Sends measured values wirelessly to logger by using a Bluetooth® wireless technology serial conversion adapter.
Power supply: 100 to 240 V AC, 50/60 Hz, 200 VA max
Dimensions: 430 x 177 x 450 mm
Weight: 14 kg.